- 157 -
order to maintain the best possible repeatability and weighing time at the
workstation. The function allows you to monitor the parameters at optional moment
and save records from carried out tests.
There are two modules:
AUTOTEST FILTER; AUTOTEST GLP.
Prior each test the balance controls level status, temperature and humidity.
AUTOTEST FILTER
Procedure consisting in loading and unloading of the internal weight for all possible
‘filter’ and ‘value release’ settings, the test requires 10 repetitions. It controls 2
parameters: ‘Repeatability’ and ‘Stabilization time’.
The test takes approximately 1 hour. Upon completion, information regarding results
is obtained. Optimum settings for particular ambient conditions are provided.
Autotest Filter enables:
to obtain the highest measurement repeatability and the shortest measurement time
with acceptable value of repeatability; test results are stored in balance memory until
it is turned off,
• test results printout,
• direct and fast selection of optimal parameters.
Upon completed autotest, a summary with results is displayed.
Information regarding filters settings is provided automatically via display of respective
pictograms:
- settings providing the fastest measurement (the shortest measurement
time).
- settings providing the most accurate measurement (the lowest possible value of
deviation for 10 measurements).
- settings providing optimal measurement (the lowest value of time and deviation
product).
- current settings for filters.
Measurement results:
*Filter type.
* <Value release> parameter value.
*Repeatability of indications expressed as standard deviation.
*Average time of measurement stabilization.