C1_A CR LF
- command understood and in progress
MASS FRAME
- command carried out, immediate response: mass value in basic measuring
unit
Response format:
1 2
3
4
5
6
7-15
16
17 18 19
20
21
S I
space
stability
marker
space
character
mass space
unit
CR LF
Switch off continuous transmission in basic measuring unit
Format:
C0 CR LF
Response options:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and carried out
Switch on continuous transmission in current measuring unit
Format:
CU1 CR LF
Response options:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
- command understood and in progress
MASS FRAME - response: mass value in current measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17 18 19
20
21
S U
I
stability
marker
space
character
mass
space
unit
CR LF
Switch off continuous transmission in current measuring unit
Format:
CU0 CR LF
Response options:
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and carried out
Set min checkweighing threshold
Format:
DH_XXXXX CR LF
, where:
_
-
space
,
XXXXX
- mass format
Response options:
DH_OK CR LF - command carried out
ES CR LF
- command not recognised (mass format incorrect)
Set max checkweighing threshold
Format:
UH_XXXXX CR LF
, where:
_
-
space
,
XXXXX
- mass format
Response options:
UH_OK CR LF - command carried out
ES CR LF
- command not recognised (mass format incorrect)
Give value of min checkweighing threshold
Format:
ODH CR LF
Response:
DH_MASA CR LF
- command carried out
Response format:
1
2
3
4-12
13
14
15
16
17
18
19
D
H
space
mass
space
unit
space
CR
LF