- 150 -
32.3.4. Test
<
Test>
parameter allowing you to verify if operation of a barcode scanner connected to a
balance is correct.
Procedure:
•
Enter
<
Barcode scanner>
submenu,
•
Select
<
Test>
parameter;
<Test>
edit box is displayed, it features ASCII and HEX
fields.
•
The code is scanned and entered to the ASCII field and HEX filed, next the test result is
displayed in the bottom part.
When:
•
<Prefix>
and
<Suffix>
declared in balance settings comply with <Prefix> and <Suffix> of
the scanned code, the test result is
<Positive>
,
•
<Prefix>
and
<Suffix>
declared in balance settings do not comply with <Prefix> and
<Suffix> of the scanned code, the test result is
<Negative>
,
32.4. Transponder Card Scanner
You can log in to the balance by means of:
−
password entered using an on-screen keyboard,
−
registered transponder card scanner.
CAUTION:
For correct ’balance - transponder card scanner’ cooperation set appropriate baud rate
value in <
Communication> submenu (by default 9600b/s).
Transponder card scanner port
Communication between the weighing device and the transponder card scanner can be
established via the following ports:
−
RS 232 (COM1)
−
RS 232 (COM2)
In order to log in an operator via transponder card scanner, first assign a number of an
already registered card to a selected operator in the database of operators.
Procedure:
•
Connect a transponder card scanner to a given port (RS 232 COM1 or RS 232 COM2),
select communication port for ‘balance-transponder card scanner’ communication.
•
Go to<
Communication> submenu and set baud rate compatible with baud rate of
the transponder card scanner (by default 9600b/s).
•
Enter operators database, go to <
Card no.> parameter and edit desired operator’s
record.
•
Upon entering editing mode in <
Card no.> parameter, a <Card no.> edit box is
opened with an on-screen keyboard.
•
Approximate the card to the transponder card scanner. The mass comparator
automatically displays number of detected card in the <Card no.> field.
•
Accept recognized number by pressing
button and return to weighing.