Value release (option disabled for the following profiles: Fast, Fast
dosing, Precision)
The parameter has been designed to determine when the stability marker
is to be displayed.
There are three different settings for value release parameter:
•
fast, fast and reliable, reliable.
CAUTION:
Both parameters, filter and value release, determine how long it takes to obtain
stable result.
Autozero Function
The function has been designed to enable automatic control and
correction of zero indication.
If the function is enabled, the subsequent measurement results are compared to
each other in constant time intervals. If the results differ less than declared
AUTOZERO range, e.g. 1 division, the balance is automatically zeroed, and the
pictograms of stable indication
and precise zero
are displayed. Active
Autozero function means, that each measurement starts with the precise zero. There
are, however, some cases when this function can be a disturbing factor for the
measuring process, e.g. very slow placing of a load on the weighing pan (load
adding, e.g. pouring, filling). In such case, the system correcting zero indication may
also correct the actual indication of a load placed on the weighing pan.
Accessible settings:
NO
- autozero function disabled
YES
- autozero function enabled
Autozero function: Dosing (function unavailable for the following
profiles: Fast, Fast dosing, Precision)
The function has been designed to set autozero to default value selected
for dosing.
Available values:
NO
- autozero operation is automatically disabled on entering dosing
mode.
YES
- autozero operation is automatically enabled on entering dosing
mode.
Last digit
The function has been designed to enable/disable visibility of the last
decimal place of measurement result. Accessible settings:
−
Always:
all digits are displayed.
−
Never:
last digit disabled.
−
When stable:
last digit displayed only when the result is stable.
Ambient Conditions
Parameter relating to ambient and environmental conditions of the
workstation. For unstable ambient conditions it is advisable to set the
parameter to value: Unstable. Default value: Stable.
Setting ‘Stable’ value results with faster operation, i.e. weighing takes
much less time than in case of setting the parameter to value: Unstable.
−
Unstable; Stable.