- 57 -
Means of operation of ACAI function:
1.
pieces quantity upon adding must increase,
2.
pieces quantity upon adding cannot increase more than twice,
3.
pieces quantity upon adding must be comprised within
±
0.3 tolerance of total value.
4.
the weighing result must be stable.
−
Minimum Reference Sample Mass
: 1 unit, 2 units, 5 units, 10 units; minimum mass
value of a single piece. If this condition is not met, the counting process will not start.
−
Result control
YES, set to print and save only those measurements which are comprised
within the Lo and Hi thresholds.
NO, set to print and save all measurements.
−
Other functions in <Settings> parameter.
To find out more about the remaining functions refer to section
15.7
of this user manual,
Additional Parameters for a Weighing Process
.
16.2. Parts Counting – Quick Access Buttons
Each working mode features automatically displayed set of default on-screen quick access
buttons. The set can be modified by assigning out-of-list quick access buttons to the on-
screen buttons. Such operation requires particular permissions.
You can select special quick access buttons that enable quick access to the determination of
single piece mass on the basis of the reference mass of the following quantity:
16.3. Setting Reference Sample Mass by Entering Mass of a Single Part
Procedure:
•
Press <
Set part mass> button, <Reference sample mass> edit box featuring an
on-screen keyboard is displayed.
•
Enter respective value and press
button for confirmation, <
Parts counting>
working mode is run with automatically set mass of a single part.
CAUTION:
If the value of entered single part mass is lower than 0.1 of the reading unit, then the
following message is displayed: <Value too low>.