- 179 -
The content of printout depends on settings of <Standard printout> parameter - <Weighing
printout template> (read section 14.5 of this user manual).
Mass printout format:
1
2
3
4 -12
13
14
15
16
17
18
stability
marker
space
character
mass
space
unit
CR
LF
Stability
marker
[space] if stable
[?] if unstable
[!] if air buoyancy compensation mode is enabled
[^] if high limit is out of range
[v] if low limit is out of range
Character
[space] for positive values
[-] for negative values
Mass
9 characters with decimal point, right justification
Unit
3 characters, left justification
Example:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF
- a printout generated,
with reference to <Weighing printout template> settings, from a weighing device upon
pressing
key:
N (measurements
quantity)
NO
Universal variable 1 … 5
NO
Date
NO
Net
NO
Time
NO
Tare
NO
Level status
NO
Gross
NO
Customer
NO
Current result
NO
Warehouse
NO
Additional unit
NO
Product
NO
Mass
Yes
Packaging
NO
Non-standard printout
NO
39. CONNECTING PERIPHERALS
The module can cooperate with the following peripheral devices:
•
computer,
•
receipt printer: KAFKA, EPSON,
•
PCL printer,
•
additional display,
•
barcode scanner,
•
an optional peripheral device operating ASCII communication protocol.
40. ERROR MESSAGES
Summary of Contents for MAS Y