- 32 -
9.4.2 Value release
Since ambient conditions at a workplace vary, it is necessary to determine the
value release parameter in a most preferable way enabling scale adaptation,
parameter options are:
FAST.+REL., FAST
or
RELIABLE
. Depending on selected
option, weighing time is either shorter or longer.
9.4.3 Autozero function
The software features autozero function (
Auto
) ensuring precise mass indication.
This function automatically controls and corrects zero indication. When Autozero is
enabled, it compares scale indications at declared time interval e.g. 1s, providing
that weighing pan is unloaded and display indication is close to zero. If results vary
less than declared AUTOZERO range e.g. one division, scale zeros automatically,
marker of stable measurement result –
and precise zero marker –
are
displayed
.
If AUTOZERO function is enabled, then each weighing process starts from precise
zero point. There are, however, some cases when this function can be a disturbing
factor for the measuring process; e.g. very slow placing of load on the weighing pan
(load pouring). Here, zero indication correction can also correct actual indication of
loaded mass.