- 138 -
x – serial number of the device (inserted in between inverted commas)
Example
:
Command
:
NB CR L
F
– give serial number
Response
:
NB_A_”1234567” – serial number of the device – ”1234567”
Value release
Format:
SS CR LF
Accessible responses:
SS_OK CR LF
- command understood, execution in progress
Command’s function is similar to function of button PRINT located on an overlay,
Internal adjustment
Format:
IC CR LF
Accessible responses:
IC_A CR LF
IC_D CR LF
- command understood, execution in progress
- adjustment completed
IC_A CR LF
IC_E CR LF
- command understood, execution in progress
- range exceeded, time limit for awaiting for a stable result
IC_I CR LF
- command understood but not accessible at this moment
Disable automatic internal adjustment of the balance
Format:
IC1 CR LF
Accessible responses:
IC1_I CR LF
- command understood but not accessible at this moment
IC1_E CR LF
- operation disabled, e.g. for verified scale
IC1_OK CR LF
- command carried out
For non-verified scales the command inhibits internal adjustment until it is enabled
via IC0 command or until the scale is turned off. The command does not modify
settings specifying adjustment start.
Enable automatic internal adjustment of the balance
Format:
IC0 CR LF
Accessible responses:
IC0_I CR LF
- command understood but not accessible at this moment
IC0_OK CR LF
- command carried out