- 53 -
The scale performs internal adjustment automatically. During adjustment process
do not load the weighing pan with any load. The message <
DO NOT TURN OFF
CALIBRATION
> is displayed in the bottom line. Once adjustment process is
completed, scale saves its result in memory and returns to weighing mode.
CAUTION!
- Press Esc button to abort adjustment process.
- If a weighing pan is loaded, message informing about an error is displayed.
Adjustment process is aborted, scale returns to the weighing mode. Adjustment
process may be repeated upon removal of the excess load.
13.3.2 External adjustment
The external adjustment shall be carried out with an external mass standard /
weigh of class F
1
.
Run an external adjustment process,
a command ordering to unload the weighing pan,
<REMOVE MASS>
(the weighing
pan must be empty), is displayed. When the weighing pan is unloaded, press
button. The scale determines mass of an empty pan, message <
CALIBRATION
> is
displayed in the bottom line. Next, message <
PUT MASS
> and weight value to be
placed on the weighing pan are displayed,
e.g.
200.000g
(depending on the scale
type). Put an external adjustment weight of displayed weight value on a pan and
press
button
.
The scale determines the weight, message <
CALIBRATION
>
is displayed in the bottom line. On completing adjustment process the scale returns
to submenu
P1.2 EXT.CALIB.
13.3.3 User adjustment
The user adjustment shall be carried out with an external mass standard / weigh of
class F
1
.
Run the user adjustment process,