- 28 -
variation at a workstation greater than 3°C). The procedure of scale adjustment is
described further down this user manual.
•
Prior to measurement, it is recommended to load the weighing pan a few
times with mass close to scale max capacity.
•
Check if unloaded scale indicates “precise zero”
and whether
measurement is stable
, if not, press
/Delete
button,
•
Press
UNITS
button,
to set a measuring unit:
•
[g], [mg], [ct], and if enabled in a service mode (non-verified scales) [kg],
[lb], [oz], [ozt], [dwt], [tlh], [tls], [tlt], [tlc] [mom], [gr], [ti], [N], [baht], [tola].
•
Place weighed object on the weighing pan and read result only on
measurement result stabilization.
•
Mass indication of a load placed on the weighing pan can be tarred
multiple number of times by pressing
→
T
←
/Insert
(pay CAUTION not to
exceed maximal capacity of a scale by applying multiple tare function).
The scale shall stay plugged to the mains in between measurement series. It is
recommended to switch off the scale display by pressing
ON/OFF
button. On
repeated
ON/OFF
button pressing, the scale is ready for operation, the
measurement can be carried out again.
9.2 Zeroing
Zeroing is a function allowing to zero mass indication. In order to zero mass
indication, press
button.
Mass indication of zero value shall be displayed together with precise zero
and stability
markers.
Zeroing process is an equivalent for determining new zero point, recognized by the
scale as precise zero. Zeroing is possible only for stable status of display
indication.