- 86 -
•
Put declared number of pieces in the container, wait until the result is
stable (the symbol
is displayed), confirm the weight by pressing
button,
•
The software automatically counts a single piece weight and enters
<
PARTS
COUNTING
> mode, number of pieces on the pan (
pcs
) is
displayed. In the bottom line, a single piece weight value is displayed (if
the option has been selected for the function <
INFORMATION
>).
CAUTION!
Remember that:
•
The total weight of all pieces put on the weighing pan must not be greater
than the device maximum weighing range.
•
The weight of a single piece cannot be lower than 0,1 reading unit. If this
condition is not met you shall see message: <Sample mass too low>.
•
While determining pieces quantity, wait for the stability pictogram
,
next confirm the quantity.
•
You can confirm the declared quantity by pressing
button only after
the stability pictogram is displayed. Otherwise, the scale will not accept
the measurement.
17.2.3 Entering Reference Sample Weight
Procedure:
•
Press F button to which the function <
ENTER SAMPLE
> is assigned, wait
to see the editing window <
ENTER SAMPLE
>
•
Use navigating buttons to enter known single piece weight value.