- 133 -
T_A CR LF
T_D CR LF
- command understood and in progress
- command carried out
T_A CR LF
T_v CR LF
- command understood and in progress
- command understood but tarring range exceeded
T_A CR LF
T_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable measurement result
T_I CR LF
- command understood but not accessible at this moment
Give tare value
Format:
OT CR LF
Accessible response:
OT_TARA CR LF
– command carried out
Response format:
1
2
3
4-12
13
14
15
16
17
18
19
O T
space
tare
space
unit
space
CR
LF
Tare
- 9 characters with right justification
Unit
- 3 characters with left justification
CAUTION!
Tare value is always given in adjustment unit.
Set tare
Format:
UT_TARA CR LF
, where
TARE
– tare value
Accessible responses:
UT_OK CR LF - command carried out
UT_I CR LF
- command understood but not accessible at this moment
ES CR LF
- command not recognised (tare format incorrect)
CAUTION:
Use dot in tare format as decimal point.
Send stable measurement result in basic measuring unit
Format:
S CR LF
Accessible responses:
S_A CR LF
S_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable measurement
result
S_I CR LF
- command understood but not accessible at this moment
S_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit