F shortcut keys
As a user you can assign weighing mode functions for quick access. The functions
shall be operated by F1, F2, F3 and F4 buttons.
List of functions to be declared for <PARTS COUNTING> mode:
NONE
no function assigned to the button
ENTER SAMPLE
function initiates an option of entering the reference
weight of a single piece
DETERMINE SAMPLE function initiates an option of automatic determination of
weight for a single piece taken from a number of samples
of known quantity
SELECT PRODUCT
function initiates an option of selecting a product from a
product database
LOG IN
function initiates an option of logging in of a selected
user
ENTER TARE
function initiates an option of entering packaging weight
SELECT TARE
function initiates an option of selecting packaging weight,
out of product database
PRINT HEADER
function initiates printout of a designed header
PRINT FOOTER
function initiates printout of a designed footer
VARIABLE 1
function initiates selection and editing of variable 1
VARIABLE 2
function initiates selection and editing of variable 2
Read “F shortcut keys” section of this user manual for procedure on assigning
functions to F buttons.
17.2.2 Reference Sample Weight determination with the use of sample of
known quantity
While determining the weight of a single piece,
ACAI
function (Automatic Accuracy
Correction) is in use.
Means of operation of ACAI function:
•
Pieces quantity (on adding) on the weighing pan has to be greater than
before.
•
Pieces quantity (on adding) on scale weighing pan must be less than twice
the amount displayed before adding parts.
•
Current quantity of pieces must be within the
±
0,3 tolerance of the total
value.
•
Measurement result has to be stable.
Procedure:
•
Put the container on the pan and tare its weight,