User Manual 

PRECISION ULTRASOUND 

BALANCE 

Model 110

 

 
 
 
 
 
 

 

E M S   P h ys i o   L td .

 

Grove Technology Park 

Downsview Road 

Wantage 

Oxfordshire OX12 9FE 

England 

Summary of Contents for 110

Page 1: ...User Manual PRECISION ULTRASOUND BALANCE Model 110 EMS Physio Ltd Grove Technology Park Downsview Road Wantage Oxfordshire OX12 9FE England ...

Page 2: ...ference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause interference to radio communications Operation of this equipment in a residential area may cause unacceptable interference to radio and TV reception requiring the operator to take wh...

Page 3: ...ese instructions must be studied before putting the unit into operation The information contained in this manual is subject to change without notice No part of this manual may be photocopied reproduced or translated into another language without the prior written consent of EMS Physio Ltd Record of Amendments ISSUE COMMENTS DATE 3 Revised 10 04 2007 4 Revised 19 10 2011 ...

Page 4: ...rmity 2 General Information 3 Record of Amendments 3 Contents 4 Warranty 5 Introduction 6 Technical Specifications 7 Unpacking 8 Assembled Balance Schematic Fig 1 9 Assembly 10 Controls and Markings 11 Operating Instructions 12 Maintenance 13 Troubleshooting 14 Parts List 15 ...

Page 5: ...lacements are carried out by the Company or its approved agents The Company will consider itself responsible for the effects on safety reliability and performance of the product only if assembly operations re adjustments modifications or repairs are carried out by persons authorised by it only if the product is used in accordance with the instructions for use only if the electrical installation of...

Page 6: ...rogrammed for direct read out in watts The unit is portable with carrying case Due to the fact that the unit is factory set if any problems arise with the unit it is highly recommended that the unit is returned for service and re calibration Environmental Considerations For best results the Precision Ultrasound Balance should only be used in an environment free from excessive air currents or vibra...

Page 7: ...city 1 litre Recommended Frequency Range 0 8 to 3 5 MHz Mains adaptor Power input AC 100 mA 240V Power output AC 500 mA 12V Other mains adaptors are available on request An optional battery pack is available All the above are typical at 23 C The balance scalefactor has been programmed assuming the temperature of the de gassed water is 23 C Readings will be approximately 2 low if the actual water t...

Page 8: ...m the carrier and the Company or its agent from whom the unit was purchased Each unit is supplied with a low voltage adaptor and cable The adaptor cable and unit all conform to international safety standards Check that the mains voltage and frequency indicated on the mains adaptor are as required If the mains supply is NOT within the specified range DO NOT CONNECT THE MAINS ADAPTOR TO YOUR SUPPLY ...

Page 9: ...ce 9 Target Assembly Clamp Support stem Support Arm Assembly Container Electronic Balance Tare Zero Button x2 Centralizing Plate Various Display ON OFF Button Figure 1 Assembled Balance Level Indicator Enter Button Select Button ...

Page 10: ...adjustable clamp for use with other styles and sizes of heads 7 The target container can now be filled with suitably degassed water To degas the water it should be boiled and left to cool in a suitable covered container see IEC 150 Sodium Sulphide at a ratio of 4g litre may be added to help degassing the water if required 8 Once the container has been filled with degassed water a suitable centrali...

Page 11: ...nce off press and hold down the On Off button Enter Button When first turned on the balance will default to show grams The Enter button will switch the display to the customised output Watts Select Button This button MUST ONLY be used to toggle between grams and watts AFTER the Enter button has been pressed Incorrect use will allow access to and possible corruption of conversion factors The EMS se...

Page 12: ...mitted from the device will hit the centre of the target cone The transducer may be held in place using a centralising plate and quick release attachment if it is an EMS device otherwise use the universal clamp supplied 5 Make sure that there are no air bubbles trapped beneath the transducer 6 Zero the balance by pressing one of the Tare buttons The balance should be re zeroed before each measurem...

Page 13: ...e container and clean them using warm soapy water Rinse thoroughly after cleaning The calibration of the Ultrasound Balance may be checked periodically using the 10 W weight provided with the balance Zero the balance Place the 10 W weight on the target frame assembly beneath the surface of the water Check that the Balance reads 10 W 0 5 W If the reading is not within the above range it is recommen...

Page 14: ...target frame is not obstructed or damaged and container is full of water 3 The power reading changes constantly Unstable ambient conditions excessive vibration or draft or a foreign object is caught between in the weighing pan frame Set up the balance in another area Remove the foreign object 4 The weight reading is obviously wrong Balance not tared before measuring Tare the balance ...

Page 15: ...de 1 A67 3 03 Absorbing Rubber Base 1 A67 3 04 Centralizing Plate 1032 large head 1 A67 3 05 Centralizing Plate 1030 head 1 A67 3 06 Centralizing Plate 1032 small head 1 A67 3 08 Centralizing Plate ML9120 1 A110 2 02 Mounting Block 1 A110 3 01 Centralizing Plate Primo SoLo large head 1 A110 4 01 Front and Rear Labels 1 L10 label EMS Physio 1 A110 5 05 Support Arm Assembly 1 B110 2 03 Stand Bracket...