Proportional
band
This is the pH (mV) span prior to the end point, at which the titrant
speed will be reduced. The speed will increasingly reduce as the end
point is approached.
For titrations where the slope of the curve is high near the end point,
the proportional band setting should be large. If overshooting of the
end point frequently occurs, increase the
Proportional band
or decrease the
Max. speed
setting. A high value for the
Pro-
portional band
will result in slow titrations.
Enter in:
Edit method > Parameters menu (end point method)
Edit reagent > Calibration parameters menu (end point method)
Range available:
-9.000 pH to +23.000 pH or -2000.0 mV to +2000.0 mV
QC (result
indicator)
Refer to "Result indicators", page 206
QC analysis
required
This message is displayed at the start of a method requiring a QC
sample.
Press
✓
and run a QC analysis.
End point no. n
Maximum
speed
Minimum
speed
Titrant delivery speed
mV or pH
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...