Titra
Lab
®
TitraLab 870 Reference Manual
Page 17
Starting up
Set up the instrument in a properly ventilated place. The power supply connector on the rear
panel must remain accessible and close to the user (2 m maximum) so that you can quickly
disconnect the cables in case of emergency.
The room temperature must be between 5 and 40°C.
The relative humidity must be between 20 and 80 %.
Connect the mains socket of the TIM870 to the mains supply using the 3-lead power cord
provided.
The TIM870 must be connected to an earthed mains socket for safety reasons.
Efficient grounding is essential for reliable measurements and security.
In the USA or Canada, use a UL listed power cord only.
Switch on the Titration Manager (
O/I
power switch set to "
I
")
The Titration Manager displays during a few seconds an identification screen (name and em-
bedded software version) followed by the Main window,
.
To a mains supply socket
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...