TitraLab 870 Reference Manual
Page 191
Reagent
calibration not
required
Message appears at the start of a sequence, when a method
sequence had been programmed with a reagent calibration.
The reagent has been reprogrammed without a calibration.
Go to Sequence/Sample stack, Edit sequence screen and remove the
reagent calibration from the sequence.
Reagent
calibration
parameters
This menu contains the general parameters concerning the reagent
calibration method (identification of the electrodes used in particular).
Titration parameters necessary to run the calibration are also pro-
grammed.
Do not forget to select the electrode and temperature sensor
created in this menu!
To access:
1.
From the Reagent window, press
4
.
2.
Select the reagent to be edited.
3.
Press
2
Edit reagent and check that the
Titre = Cali-
brate
option has been selected.
4.
Edit the calibration general parameters.
5.
Use the
LEFT/RIGHT
arrow keys to move to the last Edit reagent
display.
6.
Press
1
Calibration parameters.
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...