8.5 NaCl 0.05% standard (w/w)
With this standard, the measurements obtained with the TIM870 fit the tables published in
October 1960 by G.F. Hewitt, Atomic Energy Research Establishment, Harwell, U.K.
The
κ
T
conductivity (in µS/cm) is given a function of T temperature by the equation stated by
IEC draft 1980
(*)
:
Where:
With:
a0 = 2.11798 10
-2
a1 = 7.86011 10
-5
a2 = 1.54398 10
-7
a3 =-6.26350 10
-9
a4 = 2.27949 10
-11
Temperature range: 0 to 140°C.
(*)
"International Electrochemical Commission", January 1980, "Sub-Committee 66 /WG2".
and :
G. F. Hewit
Chemical Engineering Division
U.K.A.E.A. Research Group
Atomic Energy Research Establishment
Harwell
October 1960.
(
)
[
]
18
1
32
.
873
−
×
+
×
=
T
b
T
T
κ
4
3
2
4
3
2
1
0
T
a
T
a
T
a
T
a
a
b
T
×
+
×
+
×
+
×
+
=
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...