Titra
Lab
®
TitraLab 870 Reference Manual
Page 142
ISE Standard
addition method
- definition
Measurement method using an selective electrode (ISE) of the ion
you want to determine the concentration.
In this method, a sample measurement is performed then followed by
1 to 9 additions of a standard of known volume and concentration.
When working in the linear response zone of the ISE electrode, this
standard addition measurement method allows to determine the
sample concentration using the following equation (*):
With:
•
Csmp = sample concentration (unknown, to be calculated),
•
Vsmp = sample intial volume,
•
Ca = standard concentration,
•
Va = volume of standard added,
•
E1 = potential measured after the addition,
•
E = potential measured before the first addition,
•
S = ISE response slope at the sample temperature. This slope is
recalculated after each addition. If only one addition is performed,
the slope must be determined by calibrating the ISE electrode.
Refer to "Electrode calibration (ISE)", page 109
.
(*)
This equation does not incude the dilution factor due to the addition of a
Refer to "6. ISE measurements - ISE standard addi-
for a description of all calculations done.
In a standard addition method, the total volume of standard added
must be small in comparison with the sample volume. The accuracy of
the method depends on the following:
•
No variation of the ionic strength (add a supporting electrolyte),
•
No interfering ions present in the sample,
•
No variation of the reference electrode junction potential,
•
The number of additions performed. Higher this number is, higher
the measurements accuray will be.
With the TIM870, the additions can be:
•
programmed: the user enters the volume of one addition and the
number of additions to be performed,
•
automatic: the user enters the potential jump due to all additions
and the number of additions to be performed. The TIM870
determines automatically the volume to be added.
How to edit an ISE standard addition method?
see "ISE Standard addition method - programmation", page 143
How to run an ISE standard addition method?
Refer to "Running a method", page 216
.
Csmp
Ca
Va
×
Csmp
Va
×
(
)
E1
E
–
(
)
S
⁄
×
10
Vsmp
–
---------------------------------------------------------------------------------------
=
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...