TitraLab 870 Reference Manual
Page 235
Standard
(conductivity
standard)
Selection of the conductivity standard used to calibrate the
conductivity cell. 8 standards are available:
Figure 42: Conductivity standards available in the TIM870
The TIM870 determines the cell constant from the conductivity values
of the standard which are saved versus temperature.
Access:
1.
From the Electrode window, press
4
.
2.
Select the Conductivity type electrode to be edited.
3.
Press
2
Edit electrode and check that the
Titre = Calibrate
option has been selected.
4.
Use the
LEFT/RIGHT
arrow keys to move to the last display.
5.
Press
2
Standard line.
6.
At the ID line, select the standard used for the calibration of the
conductivity cell.
Conductivity
standard
Temperature
range
Radiometer Analytical
part no.
1 D KCl
0 to 27 °C
S51M001 (500 ml)
0.1 D KCl
0 to 50 °C
S51M002 (500 ml)
0.01 D KCl
0 to 50 °C
S51M003 (500 ml)
0.1 M KCl
0 to 36 °C
C20C250 (500 ml)
0.01 M KCl
0 to 34 °C
C20C270 (500 ml)
0.001 M KCl
0 to 30 °C
C20C280 (500 ml)
0.05 % NaCl
0 to 99.9 °C
S51M004 (500 ml)
25 µS/cm NaCl
0 to 100 °C
S51M013 (250 ml)
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...