TitraLab 870 Reference Manual
Page 215
Running a
manual dosing
To prepare a manual dosing,
Automatic add. = No
An increment is added each time key
1
is pressed.
Automatic add. = Yes
The reagent is added automatically if you press
1
. Dosing stops when
the Max. volume has been delivered or if you press
1
again.
Displaying the direct or derivative curve: press Del.
Condition: Measurements must have been performed
Stopping a manual dosing: press Stop.
The Reagent window is displayed.
Adjust the increment size (0.01 ml to 10 ml).
Select Yes to add the increments continuously
Press
1
to start dosing the reagent
(one increment is added or reagent is added
automatically increment by increment ).
Select No to add a given number of
increments.
If measurements have been selected, pH
(or mV) and/or temperature are displayed live.
With measurements: the measured
signal and/or temperature are also
= f(volume) is displayed live.
displayed. Curve measured signal
If no measurements: the volume
already dispensed is displayed.
Curve volume (in ml) = f(time)
is displayed live.
Select the curve to display.
Press
1
to quit the manual dosing.
Use the RIGHT/LEFT arrow keys to move
along the curve and display the point
coordinates shown by the cursor.
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...