Titra
Lab
®
TitraLab 870 Reference Manual
Page 206
Result ID
The name assigned by default to a method result number n is
Rn
(n=1
to 8). You can change the
Rn
name of each method result (16
alphanumerical characters maximum).
Enter in:
Edit method > Results
Edit electrode > Results
Edit reagent > Results
Refer to "Results menu", page 210
If you change a method result name (for example, from R1 to
«My result»), the equations taking this result into account
continue to use the original operand name (R1 in our
example).
Refer to "Equation formula", page 122
Result
indicators
Appear with the result at the end of a run.
•
OK: accepted by the instrument (result lies within the acceptance
limits).
•
Low: alarm, rejected by the instrument (result lies below the
acceptance limit or belw the C
o
blank concentration in the case of a
Direct ISE measurement method).
•
High: alarm, rejected by the instrument (result lies above the
acceptance limit or above 10
30
in the case of a Direct ISE or Stand-
ard addition ISE method).
•
Time max: the measurement has been accepted at the end of the
Acceptance delay.
•
QC: the user has bypassed a QC sample analysis demand.
•
Man. stop: result has been obtained from a titration that was
stopped manually (
Del
key).
•
0IP: no inflection points found.
•
IP>1: more than one inflection point has been found in the min/max
ordinate window,
see "Min. ordinate Max. ordinate", page 155
•
Vstd>max: In an ISE Standard addition method with automatic
determination of the volumes, a Vstd>max note is added to the
result if the cumulate volumes of the standard additions is higher
than 50 % of the sample test quantity.
Refer to "ISE Standard addition method - programmation", page
143
You can reject a result that was accepted by the instrument.
You can accept a result that was rejected by the instrument
(Supervisor users only).
Refer to "Result accepted (Yes/No)", page 205
.
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...