User Manual 3162A/B
Maintenance and Performance Checks 5-8
4.
Configure Channel B using the following commands:
Remote Commands: :INST:SEL B
:VOLT 10e-3
:VOLT:OFFS 2.495
:OUTP ON
5.
Connect the B channel output to the DMM input. Terminate the
output with a 50
Ω
feedthrough termination. Set the B channel offset
and verify DMM reading as follows:
B Channel Offset Setting
DMM Reading
+2.495 V
+2.435 V to +2.555 V
+1.000 V
+0.980 V to +1.020 V
-1.000 V
-0.980 V to -1.020 V
-2.495 V
-2.435 V to –2.555 V