User Manual 3162A/B
Maintenance and Performance Checks 5-20
Amplitude
Adjustment
Equipment:
DMM, 50 feedthrough termination
1.
Configure the Model 3162B as follows:
Function:
Sine
Frequency: 1 kHz
Amplitude:
5 V
Remote Commands: :RES
:FREQ 1e3
:VOLT 5
:OUTP ON
Channel A Adjustment
2.
Connect the Model 3162B output to the DMM input through the 50
Ω
feedthrough termination. Set the DMM to AC volt measurements and 2
V range.
3.
Adjust R34 on the engine board A for DMM reading of 1.7677V, ±15
mV.
Channel B Adjustment
4.
Change the Model 3162B settings using the following remote
commands:
:INST:SEL B
:VOLT 5
:OUTP ON
5.
Connect the Model 3162B output to the DMM input through the 50
Ω
feedthrough termination. Set the DMM to AC volt measurements and 2
V range.
6.
Adjust R78 on the engine board B for DMM reading of 1.7677V, ±15
mV.