User Manual 3162A/B
Maintenance and Performance Checks 5-7
4.
Configure Channel B using the following commands:
Remote Commands: :INST:SEL B
:OUTP ON
:VOLT 5
5.
Connect the B channel output to the DMM input. Terminate the
output with a 50
Ω
feedthrough termination. Set the amplitude and
verify DMM reading as follows:
B Channel Setting
DMM Reading
5.000 V
1.725 V – 1.809 V
500.0 mV
170.0 mV – 183.4 mV
50.0 mV
15.50 mV – 19.84 mV
DC Offset
Characteristics
Accuracy Specifications:
± (2% of r10 mV)
Equipment:
DMM, 50 feedthrough termination
1.
Configure the Model 3162B as follows:
Function:
Sine
Frequency:
1.000 MHz
Amplitude:
10 mV
Remote Commands:
:RES
:FREQ 1e6
:VOLT 10e-3
:VOLT:OFFS 2.495
:OUTP ON
2.
Set the DMM to DCV measurements.
3.
Connect the A channel output to the DMM input. Terminate the
output with a 50
Ω
feedthrough termination. Set the A channel offset
and verify the DMM reading as follows:
A Channel Offset Setting
DMM Reading
+2.495 V
+2.435 V to +2.555 V
+1.000 V
+0.980 V to +1.020 V
-1.000 V
-0.980 V to -1.020 V
-2.495 V
-2.435 V to –2.555 V