Communication Protocol
CubiScan 125 Command Set
CubiScan 125 Service Manual
7 0
* This field may contain underscores, dashes, or tildes indicating an under, unstable, or over
error condition, respectively. Leading spaces (20h) will be used when the actual data does not fill
the entire field.
Scale Calibration
This function is required when the load cell is replaced, when a new control is installed, or for
routine calibration. This command causes the CubiScan 125 to reply with a scale calibration code.
This command must be issued THREE times to complete scale calibration. Scale calibration
codes are defined below.
7
1
Carriage Return
Control
(CR)
0Dh
8
1
Line Feed
Control
(LF)
0Ah
Pos Len
Description
Type
Range
ASCII
Command Format
1
1
Start of Text
Control
(STX)
02h
2
1
Scale Calibration
Alpha
(S)
53h
3
6
Test Weight Value
Numeric
050.00-100.00
9
1
End of Text
Control
(ETX)
03h
10
1
Carriage Return
Control
(CR)
0Dh
11
1
Line Feed
Control
(LF)
0Ah
Acknowledge Format
1
1
Start of Text
Control
(STX)
02h
2
1
Scale Calibration
Alpha
(S)
53h
3
1
Acknowledge
Alpha
(A)
41h
4
2
Identifier
Numeric
(00-99)
6
1
End of Text
Control
(ETX)
03h
7
1
Carriage Return
Control
(CR)
0Dh
8
1
Line Feed
Control
(LF)
0Ah
Negative Acknowledge Format
1
1
Start of Text
Control
(STX)
02h
2
1
Scale Calibration
Alpha
(S)
53h
3
1
Neg. Acknowledge
Alpha
(N)
4Eh
4
1
End of Text
Control
(ETX)
03h
5
1
Carriage Return
Control
(CR)
0Dh
6
1
Line Feed
Control
(LF)
0Ah
Pos Len
Description
Type
Range
ASCII