Communication Protocol
CubiScan 125 Command Set
CubiScan 125 Service Manual
6 8
* This field may contain underscores, dashes, or tildes indicating an under, unstable, or over
error condition, respectively. Leading spaces (20h) will be used when the actual data does not fill
the entire field.
Gate Measure
The CubiScan 125 will automatically send measurement data upon completion of a gate
measurement.
51
1
Wgt./Dim.Wgt Unit
Alpha
(E/M)
45h or 4Dh
52
1
Comma
Alpha
(,)
2Ch
53
1
Factor Identifier
Alpha
(F)
46h
54
4
Factor
Numeric
0000-9999
58
1
Comma
Alpha
(,)
2Ch
59
1
Domestic/Int'l. Unit
Alpha
(D/I)
44h or 49h
60
1
End of Text
Control
(ETX)
03h
61
1
Carriage Return
Control
(CR)
0Dh
62
1
Line Feed
Control
(LF)
0Ah
Negative Acknowledge Format
1
1
Start of Text
Control
(STX)
02h
2
1
Measure Command
Alpha
(M)
4Dh
3
1
Neg. Acknowledge
Alpha
(N)
4Eh
4
1
CubiScan/Host
Originated
Alpha
(C/H)
43h or 48h
5
1
Corner Sensor / Measure
/ Zero Error
Alpha
(C/M/Z)
43h or 4Dh
6
1
End of Text
Control
(ETX)
03h
7
1
Carriage Return
Control
(CR)
0Dh
8
1
Line Feed
Control
(LF)
0Ah
Pos Len
Description
Type
Range
ASCII
Acknowledge Format
1
1
Start of Text
Control
(STX)
02h
2
1
Measure Command
Alpha
(m)
6Dh
3
1
Acknowledge
Alpha
(A)
41h
4
1
CubiScan/Host
Originated
Alpha
(C/H)
43h or 48h
5
6
City Code
Alpha
000000-ZZZZZ
Z
Pos Len
Description
Type
Range
ASCII