Scale Calibration
Scale Calibration
CubiScan 125 Service Manual
1 7
6. Make sure the CubiScan 125 platform is clear of all objects and tap
[Next]
to continue. The
following screen is displayed.
7. Place the test weight onto the glass platform (roughly in the center) of the CubiScan 125.
Allow the scale to completely stabilize, and tap
[Next]
.
8. When the CubiScan 125 is done calibrating, tap
[Prev]
twice to return to the menu screen.
Figure 8
Second Scale Calibration Screen
Figure 9
Third Scale Calibration Screen