44
| Software
|
Ultra
Version M 22-April-2012
Feature
Operation
Command
Response
format
Description
Status
Commands
(continued)
Q-Switch Interlock Byte, query
Example: IQS
IQS 01 00 00 00
indicates coolant is too cold
IQ
IQS ab cd ef gh
The response string contains ASCII-encoded 1’s or 0’s
indicating that either a fault or okay condition exists.
a:
“1” if in the 8-second forced delay after
flashlamp starts
0 if not
b:
“1” if coolant temperature is too low
0 if not
c:
“1” if Q-Switch timeout
0 if not
d:
“1” if shutter is closed
0 if not
e:
not used
0 if not
f:
not used
0 if not
g:
not used
0 if not
h:
not used
0 if not
First Interlock
Example: IF
IF2 00 00 01 00
IF
??? ab cd ef gh
Queries the interlock that caused a laser shutdown. The
response is that of IF1, IF2, IF3 or IQ depending on which
one caused the shutdown. In the case of multiple causes,
priority is from IF1 down to IQ.
Temperature
Commands
Cooling Group Temperature
Example: CGT
temp. CG 45 d
indicating the coolant temperature is 45°C
CGT
temp. CG nn d
Queries the temperature of the coolant in degrees Celsius
Harmonic Generator Temperature
Example: HGT
temp. SHG 75
d
indicates the harmonic generator temperature is
75°C
HGT
temp. SHG nn d
Queries the temperature of the harmonic generator in
degrees Celsius
Charger/Simmer Temperature
Example: CST
temp. CS 55 d
indicates the charger/simmer temperature is 55°C
CST
temp. CS nn d
Queries the temperature of the charger/simmer controller in
degrees Celsius