| Troubleshooting
|
Ultra
Version M 22-April-2012
Interlock
Value
ICE Front Panel
Message
Possible Cause
Solution/Suggestion
Description
IF3 00010000
low frequency
The external trigger
input frequency is
less than frequency
minimum value.
Decrease the frequency of the external
signal generator
or
increase the setpoint for frequency.
Contact Quantel Customer Service for
details on disabling frequency checking.
When operating in External flashlamp
triggering mode, the ICE450 compares the
input trigger frequency to the programmed
frequency as viewed on the Remote Box.
An interlock condition occurs if they do not
match.
IF3 00001000
high frequency
The external trigger
input frequency is more
than the maximum
frequency value.
Either increase the frequency of the
external signal generator
or
decrease the set-point frequency.
Contact Quantel Customer Service for
details on disabling frequency checking.
When operating in External flashlamp
triggering mode, the ICE450 compares the
input trigger frequency to the programmed
frequency on the Remote Box. If they do
not match, an interlock condition occurs.
IF3 00000100
Capacitor not
discharged
The flashlamp capacitor
did not completely
discharge after the
flashlamp turned off.
CAUTION:
This
is due to a fault
in the charging
system and may
be hazardous.
You should not attempt to repair this
problem. Contact Quantel Customer
Service for repair.
The system controller monitors the voltage
on the PFN capacitor. The system displays
this interlock if there is an unsafe residual
voltage remaining on the PFN capacitor.
IF3 00000010
simmer disabled
time out delay
expired
The simmer timeout has
expired.
You should not attempt to repair this
problem. Contact Quantel Customer
Service for repair.
You should not attempt to repair this
problem. Contact Quantel Customer
Service for repair.
NOTE: To clear an interlock message, press the “Cursor
Up” or “Cursor Down” button on the Remote Box.