MAINTENANCE
CAUTION:
Inadequate cooling system
maintenance may result in coolant contamination
and/or system damage.
1.0 Scheduled Maintenance
Perform the following maintenance procedures on a
regular schedule:
Turn the Key Switch ON to operate the pump and
circulate coolant for at least 30 minutes every
month when the laser is not in use.
Inspect the coolant level in the reservoir through
the reservoir-level window on the ICE450.
Replace the Deionizing cartridge every 6 months
and each time the flashlamp is replaced to
maintain coolant integrity.
Replace the flashlamp after every 50 million shots.
2.0 Draining the ICE450
CAUTION:
Remove ALL coolant.
Coolant trapped within the Power
Supply may freeze and cause
irreversible damage to the internal
components.
Note:
For the standard ICE450, as
opposed to the rack mount
version, you must position the
ICE450 onto its side with the
Fill/Drain port down (Vent up).
Note:
It is useful to have a second person
assist in draining the ICE450. You
will need a suitable container to
hold the drained liquid.
Use the following procedure to drain the
system:
1. Turn the Key Switch OFF.
2. Position the ICE450 on its side with
the Fill/Drain port down (Vent port up).
3. Turn the Key Switch ON.
4. Use the Remote Box Pump menu to
turn the Pump OFF.
5. Disconnect the coolant lines from
the back of the ICE450 and from the
Laser Head.
6. Connect the supplied Coolant Loop
between the blue and red fittings on
the ICE450 rear panel.
Disconnect the coolant
lines from the blue and
red fittings on the rear of
the ICE450
Connect
Coolant
Loop
Coolant
Loop
Fill/Drain
tube
Vent tube