36
| Software
|
Ultra
Version M 22-April-2012
2.0 Serial Command Reference
Feature
Operation
Command
Response
format
Description
Flashlamp
Commands
Fire, Internal Sync
Example: A
fire auto
A
state string
Sets the flashlamp into “Internal” sync mode and to start
flashing. The response to this command is the state string
(see the description for the “ST” command located in the
Status Commands section). If the flashlamp is able to start
flashing, the response will be “fire auto”. If not, the response
will indicate the state.
Fire, External Sync
Example: E
fire ext
E
state string
Sets the flashlamp into “External” sync mode and to start
flashing. The response to this command is the state string. If
the flashlamp is able to start flashing, the response will be
“fire ext”. If not, the response will indicate the state.
Simmer
Example: M
simmer
M
state string
Sets the flashlamp into Simmer mode. The response to
this command is the state string. If the flashlamp is able to
start simmering, the response will be “simmer”. If not, the
response will indicate the state.
Stop Flash
Example: S
standby
S
state string
Stops the flashlamp from flashing and turns off the simmer.
The response to this command is the state string, and it
should be “standby”.
indicates entering the command. This can be done by sending a carriage return
character, CR (‘\r’, 0x0D) and line feed character, LF (‘\n’, 0x0A), in that order.