6.0 Remote Box Navigation
Use the LCD display to select Ultra Laser System functions.
The Cursor:
The cursor “
>
” is used to navigate the
Remote Box menu.
Moving the Cursor:
Use the “Cursor Up/Down” buttons
to move the cursor to the next line in the current menu.
The LCD screen shows only four items. Move the cursor to
scroll to items that are not visible on the display.
Selecting a Menu Item:
There are three menu levels.
Use the “Enter Menu” button to access sub-menus (
).
With the cursor in front of an item, press the “Enter Menu”
button to show the corresponding menu. “Enter Menu”
lines have a small arrow on the right side in the diagram.
Example: position the cursor at the
Flashlamp
item and
press the “Enter Menu” button.
Note:
Use the “Enter Menu” button to initiate action items
such as
Save
, to store a configuration.
Change Values/Settings:
Use the “Increase/Decrease
Value” buttons to change a setting. Examples: To change
the
QS mode
, position the cursor at this item, then press
the Increase/Decrease Value button to switch between
AUTO, BURST and SCAN. To decrease the
FLQS dly
, with
the cursor at this item, press the “Decrease Value” button.
In the diagram, items allowing value adjustment or mode
selection options are shown highlighted.
See
Remote Box Menus, Detailed Descriptions
.
Figure 10
BASIC MENU Structure For the Remote Box
LCD screen
Cursor UP
Cursor DOWN
Increase Value
Decrease Value
Enter Menu
button
Main Menu
configuration
1
>Flashlamp
Q-Switch
System Info
Save
1
QS Menu
Main menu
QS sync
INT
FLQS dly
120
µ
s
>QS mode
AUTO
Ramp
System Info Menu
Main Menu
>Temp
34.1°C
serial link OFF
Time-out
OFF
ICE450/BRIO
version 3.05
option HG OFF
Interlocks
Pump
ON
Flash Menu
Main menu
>flash sync INT
ct
0000.001.368
cu
0000.001.368
voltage
550V
energy
06.77J
freq.
20.00Hz
power
132W
User FL Counter Menu
Flash Menu
>reset counter
cu
0000.005.242
QS Mode/Auto Menu
QS Menu
>QS mode
AUTO
QS pulse
F/02
QS Mode/Burst Menu
QS Menu
>QS mode
BURST
QS pulses 009
QS Mode/Scan Menu
QS Menu
>QS mode
SCAN
scans
30
active
5
passive
10
QS Ramp Menu
QS Menu
>ramp ON
000
r step
10
µ
s
QS at run OFF
Temp Menu (read only)
System Menu
>CG temp
34.1°C
HG temp
37.1°C
CS temp
43.2°C
Tmin fl
30°C
Tmax fl
55°C
Tmin qs
35°C
heater
44%
flow
1.805 lpm
level
OK
Time out Menu
System Menu
>time sim. 00:00
time fl
00:00
time qs
00:00
Interlocks Menu
System Menu
>IF1
00000000
IF2
00000000
IF3
00000000
IQS
00000000
Pump Menu
System Menu
>mode
normal
pump
ON
level
OK
flow
3.634 lpm
This arrow indicates
a submenu item.
This arrow indicates
return to the previous
menu.
Use the Increase/Decrease
Value buttons to toggle
values (in this case
between INT internal and
EXT external.) See Remote
Box Menus, Detailed De-
scriptions for more informa-
tion on command options.
Use the Cursor
UP/DOWN buttons
to scroll and show
items not on screen.
Cursor in