78
| Troubleshooting
|
Ultra
Version M 22-April-2012
1.0 General Information
No
internal
parts of the ICE450 are user serviceable. The only items on an Ultra laser system
that may be serviced by the user are:
•
DI cartridge, see
Replacing the DI Cartridge on page 72
•
Flashlamp, see
Flashlamp Replacement on page 74
•
10A Fuses see
Replacing the ICE450 Fuses on page 73
All other service or repair of the ICE450, Laser Head or Laser Head optical alignment issues
require a qualified Quantel technician or trained Field Service Representative.
Basic troubleshooting can help you resolve:
•
security system faults, such as external safety interlocks
•
connection issues
If a failure occurs at start-up or during operation, use the Key Switch to shut the laser OFF.
Then use the troubleshooting methods presented next to resolve the issue.
2.0 Interlocks
The ICE450 constantly monitors the system for potentially hazardous conditions. If any of
these conditions exist, the laser operation ceases and the Remote Box shows a message
indicating the interlock condition. The “IF” command can also be used to query interlock
conditions via the serial interface (See details in
NOTE:
To clear an interlock message, press the “Cursor Up” or “Cursor Down” button on the
Remote Box.
TROUBLESHOOTING
The ICE is designed to control the laser and warn the
user of problems that may occur. The microprocessor-
based system monitors the laser system and
automatically shuts down if a fault occurs. Software
limits have been factory selected to protect the laser
system against damage. The following section may
help you correct or identify the problem before calling
Quantel Customer Service.
Contact Us
QUANTEL HEADQUARTERS (Worldwide)
2 bis avenue du Pacifique – Z.A. de Courtaboeuf
BP 23 – 91941 Les Ulis Cedex – France
Phone: +33-1-6929-1700
Fax : +33-1-6929-1729
[email protected]
QUANTEL USA (North America)
601 Haggerty Lane
Bozeman, MT 59715 USA
Service toll free: +1-800-914-8216
Fax: +1-406-522-2007
[email protected]
.