General Description
PyroMark Q96 ID User Manual 01/2016
3-7
3.8
Controls and indicators
3.8.1
Front panel
The standby switch is located on the lower left-hand side of
the front panel. Three indicator lights can be found in the
middle of the front panel:
Top light: Power (green). When illuminated, this indicates
the instrument has been connected to the mains
electricity supply and is receiving power
Middle light: Busy (yellow). When this light is illuminated
or flashing, a run is in progress and the lid should not be
opened
Bottom light: Information (yellow). If this light is
illuminated during a run, the instrument has lost contact
with the operator’s computer
During the start-up sequence:
All the lights illuminate
The instrument software loads (takes 1–3 minutes) and
the “Busy” light goes off and the “Information” light
starts flashing. This indicates that the instrument is ready
to receive instructions from the operator’s computer
When an application is selected and “Run” is clicked in
the “Setup Run” window, the flashing “Information” light
will turn off and the “Busy” light will illuminate
3.8.2
Rear panel
The following connectors are located on the rear panel:
Mains power input connector: Should be connected to a
grounded mains power outlet
Keyboard: Keyboard interface for service purposes. This
should only be used by a certified QIAGEN Field Service
Specialist
Mouse: Mouse interface for service purposes. This should
only be used by a certified QIAGEN Field Service
Specialist
Net: Ethernet network connector
Summary of Contents for PyroMark Q96 ID
Page 1: ...Version 5 January 2016 Sample Assay Technologies PyroMark Q96 ID User Manual...
Page 2: ......
Page 8: ...Contents Contents 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 14: ...Safety Information 1 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 78: ...Maintenance 7 12 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 96: ...Appendix A A 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 106: ...Appendix C C 2 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 114: ...Appendix D D 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 122: ...Appendix E E 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 125: ......