Glossary
PyroMark Q96 ID User Manual 01/2016
8-1
8
Troubleshooting
If you need to contact QIAGEN Technical Services about an
error, note down the steps leading to the error and any
information given in any dialog boxes. This will help the
QIAGEN Instrument Service Specialist in solving the problem.
When calling QIAGEN Technical Service about errors, please
have the following information ready:
Instrument serial number, type, and version
Date of last maintenance performed
Error code (if applicable)
Time point when the error occurred for the first time
Frequency of error occurrence (i.e., intermittent or
persistent error)
Photo of error, if possible
Take the following action before contacting QIAGEN
Technical Services.
1.
Check important parameters in the “Run Overview” in
the “Instrument Status” tab to assess if the system was
working properly during the run.
2.
Consult the Troubleshooting sections below.
8.1
Unexpected results
Comments and suggestions
a) No result in any
wells of the plate
Check the reagent cartridge and ensure that it is
correctly filled with reagents and/or nucleotides.
Check that the needles are not damaged or
obstructed and that the reagent cartridge cover is
closed.
Insufficient sample or incorrectly conditioned
sample (e.g., too much dsDNA or salt present).
Follow sample preparation instructions carefully
(see Section 5.4).
Check the PCR samples using a gel technique to
confirm there is only one specific band.
Summary of Contents for PyroMark Q96 ID
Page 1: ...Version 5 January 2016 Sample Assay Technologies PyroMark Q96 ID User Manual...
Page 2: ......
Page 8: ...Contents Contents 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 14: ...Safety Information 1 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 78: ...Maintenance 7 12 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 96: ...Appendix A A 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 106: ...Appendix C C 2 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 114: ...Appendix D D 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 122: ...Appendix E E 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 125: ......