Operating Procedures
PyroMark Q96 ID User Manual 01/2016
5-5
2.
Right-click an existing instrument parameter file listed in
the upper pane and select “Duplicate” from the context
menu.
3.
Enter a new file name (e.g., “Instrument parameters
Code 0005”).
4.
Go to the Web page cited above and open the file
“Managing instrument parameters for the PyroMark Q96
ID”, found under the User Support tab. In the provided
table, find the dispensing pressures and pulse time
settings that correspond to the Code number printed on
the PyroMark Q96 Cartridge to be used.
5.
Enter the dispensing pressures and pulse time into the
corresponding fields of the “Instrument Parameters”
dialog box.
6.
Save the new instrument parameter file.
7.
When creating a new Run Setup, select this new
instrument parameter file from the list of instrument
parameter files.
Summary of Contents for PyroMark Q96 ID
Page 1: ...Version 5 January 2016 Sample Assay Technologies PyroMark Q96 ID User Manual...
Page 2: ......
Page 8: ...Contents Contents 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 14: ...Safety Information 1 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 78: ...Maintenance 7 12 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 96: ...Appendix A A 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 106: ...Appendix C C 2 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 114: ...Appendix D D 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 122: ...Appendix E E 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Page 125: ......