91
11.3
Advanced Auto-Mode
NOTE:
This section is only for advanced users!
For advanced applications, the auto-mode can be configured even
more flexible as described in the preceding chapter. First, it is
possible to define an independent sample interval for each oxygen
channel; e.g. channel 1 is measured every 10 s, while channel 2 is
measured
only
every
10 min.
Second,
optionally
the
measurements of each channel can be triggered externally by
using the trigger input of the extension port (see appendix 12.2).
The advanced auto-mode can be configured by opening the
Settings
window and clicking there on the
Advanced Auto-Mode
button in the
Options
tab. A separate window will open.
Three options can be selected. The first option
Disable Auto-Mode
disables any measurement in the auto-mode for the respective
channel (but it does not disable the auto-mode in general!). For the
second option
Interval Auto-Mode
, the sample
Interval (s)
can be
adjusted for each channel independently. If the last option
Triggered Auto-Mode
is chosen, the measurement of the
respective channel is only performed during the auto-mode if a