37
NOTE
: The Charts are automatically cleared in the panels of the
Main Window
after the
Settings
have been modified. Re-
adjustments in the
Settings
might require also a recalibration of
the sensor(s).
If the changes of the Settings require a sensor recalibration, a
warning
Not Calibrated
appears right-hand side of the
Calibrate
button in the corresponding channel panel of the Main Window.
6.2.1
Basic Settings
The
Sensor Settings
can be adjusted in a
Basic
or an
Advanced
mode. The first-time user is advised to work with the
Basic
Settings
. Please ensure that the correct sensor code attached to
the sensor has been entered in the field
Sensor Code
.
For contactless sensors (sensor spots, flow-through cells,
respiration vials; sensor type:
S, W, T) and for robust probes
(sensor type: X), additionally the
Fiber
Length (m)
of the connected optical
fiber (e.g.
SPFIB
) or of the connected
robust probe (e.g.
OXROB10
) must be
entered (for interested users: the
entered fiber length is used for an
automatic background compensation;
refer to chapter 9.2 for more details).
The
Measuring Mode
can be adjusted gradually between low drift
(1) and low noise (5) of the sensor signal by moving the arrow with
the mouse along the scale, thereby changing the oxygen
measuring time. An intermediate mode (3) is default.
NOTE
: Ensure that the correct sensor code has been entered. If the
sensor is not yet calibrated, a warning
Not Calibrated
is displayed
in the Main Window.