go to the
Settings
, enter the correct
Sensor Code
and repeat the
calibration.
Then the air temperature
needs to be determined. Either a
Fixed
Temperature
is adjusted manually or the temperature is read from
the external temperature sensor by selecting
External
Temperature Sensor
.
Wait now until the sensor reading is stable by observing the graph
and the numerical display of the oxygen sensor reading. If
External
Temperature Sensor
was selected, ensure also stable temperature
readings indicated at
Temperature Sensor (°C)
. Note that the
button
Set Air
will be highlighted as soon as the oxygen readings
are within the expected range for the connected sensor type.
If all readings have reached their steady state, click on
Set Air
, and
the actual oxygen sensor reading is taken for the air
calibration. If the oxygen reading seems to be out of
the expected range, a warning will be shown offering
the possibility to repeat the calibration. A completed
calibration is indicated by the green indicator
Calibrated at Air
.
Finally, click on
Finish
for returning to the main window.