42
After activation of a temperature sensor in the oxygen channels
(see 6.2.3), the measured temperature by the external or internal
temperature sensor is displayed in the panel(s) of the
corresponding oxygen sensors with activated automatic
temperature compensation (right-hand side y-axis). It is also
displayed in the
Overview
panel (see chapter 0) of the main
window and saved into the data file.
The temperature is displayed in units of degree Celsius (°C).
The
Temperature Measuring Time
(default 300 ms; only multiples
of 100ms possible) of the
External Temperature Sensor
can be
optionally (a) increased in order to reduce the noise of the
temperature measurement or (b) decreased in order to achieve
higher sampling rates.
A
Manual Offset
of the External Temperature Sensor might be
entered for possible recalibration of the temperature sensor
(default: 0).
Please note that each
FireStingO2
meter provides only a single
port for a temperature sensor. For
Automatic Temperature
Compensation
(see chapter 6.2) of >1 oxygen sensor, all sensors
with automatic temperature compensation have to measure at the
same temperature condition, as determined with the connected
external temperature sensor (see also chapter 10). Alternatively, if
the temperature extension module
TeX4
is coupled to the
FireStingO2
, then automatic temperature compensation can be
performed for each channel independently using the temperature
sensor connected to the corresponding port of the
TeX4
.