NOTE
: If using retractable needle-type sensors (e.g.
OXR50
,
OXR230
,
OXR430
), it is important that the sensor tip is extended to
position 2 or 3 (see chapter 5.1) when the calibration value is taken.
7.1.6
Calibration Mode: Custom Mode
The custom calibration mode is selected by clicking on the button
Custom Mode
. This mode allows the user to combine freely all
possible calibration types for the air calibration and the 0%
calibration. The air calibration type can be selected by clicking on
the “Air Calibration Selector”. The 0% calibration type can be
selected by clicking on the “0% Calibration Selector”.
The following air calibration types can be chosen:
Factory Air
Calibration
(refer to chapter 0)
Air with 100% Humidity or Air Saturated Water
(refer to
chapter 7.1.3)
Ambient Air
(refer to chapter 7.1.2)
Custom Calibration
(more details below)