During the calibration of oxygen sensors, there are two possibilities
to take the humidity into account:
(1)
The relative humidity and the temperature of the
ambient air must be determined during the calibration.
The
Pyro Oxygen Logger
software calculates then
automatically the real oxygen level under these
conditions.
(2)
The calibration standard is prepared in a closed vessel
either filled with water or partly filled with e.g. wet
cotton wool or a wet sponge. This ensures a constant
humidity of 100%RH and there is no need to measure
the humidity.
Option (1) is utilized in the calibration standard “Ambient Air” (see
section 8.1.1), whereby option (2) is utilized for the calibration
standards “Water-Vapor Saturated Air” (see section 8.1.2) and “Air
Saturated Water” (see section 8.1.3).
Another parameter even more important for the air calibration
standard is the atmospheric pressure. The principle parameter
measured by oxygen sensors is not the partial volume (i.e. “%O2”),
but the partial oxygen pressure (i.e. “mbar”) (see also appendix
12.6). So an oxygen level of e.g. 20.7%O2 (determined as described
above by a given humidity and temperature) is converted internally
by the
Pyro Oxygen Logger
software into a partial pressure of
oxygen essentially by multiplying the relative oxygen level with the
atmospheric pressure of e.g. 990 mbar (see chapter 12.6):
0.207 x 990 mbar = 205 mbar
giving a partial oxygen pressure of e.g. 205 mbar. This is the
essential calibration value used internally by the
Pyro Oxygen
Logger
software. The atmospheric pressure can be influenced 1) by
weather changes (e.g. varying between ca. 990 and 1030 at sea
level) and 2) by the elevation above sea level (e.g. at 1000 m