12.3
Troubleshooting
How to respond to the warnings shown in the
Pyro Oxygen
Logger
:
Signal Too High
Too much ambient light exposed to the sensor, or amplification is
too high, or LED intensity is too high:
→
darken the surrounding
→
and/or decrease
Amplification
in the
Advanced
settings
→
and/or decrease
LED Intensity
in the
Advanced
settings
Low Signal
Sensor signal is too low:
→
check whether the sensor cable is connected
→
increase
Amplification
in the
Advanced
settings
→
and/or increase
LED Intensity
in the
Advanced
settings
→
replace sensor, the tip might be broken/bleached
Bad Reference
Internal problem of the electronics
→
contact
PyroScience