66
7.2.1
Calibration Mode: Factory (1. Generation
FireSting
)
NOTE
: The
Factory Calibration
(only for rough measurements
and testing purposes)
is only possible if the
correct Sensor Code
has been entered in the
Settings
(see chapter 6.2).
If the calibration mode
Factory Calibration
is chosen, ensure that
the correct sensor code has been entered in the
Settings
of the
corresponding channel (as displayed in 2.
Adjust Calibration
Conditions
of the
Oxygen Sensor Calibration
window). If the
sensor code displayed is not correct, click on
Finish
, go to the
Settings
, enter the correct
Sensor Code
and repeat the
Factory
Calibration
.
After clicking on
Finish
the factory calibration is completed,
thereby returning to the main window of the corresponding
channel.
7.2.2
Calibration Mode: 1-Point (1. Generation
FireSting
)
NOTE
: The calibration mode
1-Point
is only possible if the
correct
Sensor Code
has been entered in the
Settings
(see chapter 6.2).
The calibration mode 1-point is selected by clicking on the button
1-Point
.
This mode uses a manual calibration in an air calibration
standard for adjusting the air calibration value. The 0% calibration
value is taken from the
Sensor Code
. The preparation of
appropriate air calibration standards is described in chapter 8.1.
Depending on the air calibration standard used, select
Water (DO)
for dissolved oxygen or
Gas Phase
.
The air calibration standard (see also chapter 8.1) can be:
ambient air of known humidity,
water-vapor saturated air or
air saturated water (100% air saturation).