46
6.4
Raw Data Window
The Raw Data Window is mostly intended for trouble shooting and
advanced users. During normal measurements it is in general not
needed. After clicking on the
Raw Data
button in the
Main
Window
(see chapter 6.1) the following
Oxygen Sensor Raw Data
window opens:
NOTE:
While the Raw Data window is opened, all raw values are
also saved into the data file in additional columns behind the
standard data columns.
The panels of each oxygen channel (
Chan 1-4
) show the phase shift
as "delta phi" (
dphi
, in °).
dphi
is the actual measured raw
value which is used for the
internal calculation of the
oxygen concentration
Oxygen
(µM)
,
Oxygen
partial pressure
(hPa)
,
Oxygen
in % air
saturation
(% air sat)
and
Oxygen
(
% O2)
(see also
chapter 12.4).
The
Signal Intensity
(in mV)
gives a measure of the quality
of the oxygen measurement,
which is also displayed in the
horizontal bar indicator in the main window (see chapter 6.1).
Ambient Light
(in mV)
gives a measure of the ambient light
entering the sensor from outside. At too high ambient light levels
the detector of the
FireStingO2
might get saturated giving the
warning
Signal too high
in this window and in the warning display
of the main window (see chapter 6.1).