Figure 56. Remove the Aperture.
5.
Soak the Injector Tip Holder and the Aperture in a 70% ethanol solution for 30 minutes to dissolve reagent
residue. Rinse with deionized water.
6.
Allow the Injector Tip Holder and Aperture to air dry completely before proceeding. Alternatively, blot dry
with a lint-free, Kimwipes
®
tissue.
7.
Return the Aperture to the Optical Head. Align the Aperture with the Optical Head and gently guide the
Aperture onto the Optical Head (Figure 57).
Figure 57. Replace the Aperture.
8.
Return the Injector Tip Holder to the Optical Head and tighten the hand screws. Position the holder just
above the front pins on the Optical Head. Push the holder toward the right to lock the holder securely to the
right pin of the Optical Head. Then push down and lock the holder on the left pin (Figure 58).