Self Test
Each time the instrument is powered “ON”, it automatically performs a self test to check the motion of all axes.
This button allows users to manually initiate a self test.
Tools
Tools may be accessed from the
Tools
icon in the menu bar (shown below) or from the Settings
Tools window.
13623T
A
Figure 40. Tools icon.
Injectors
Dual Injectors are an optional accessory for the GloMax
®
Navigator Instrument.
Injector System Specifications
Number of Injectors
Two Injectors
Dispense Volume Range
5–200µl in 1µl increments
Injection Speed
20–500µl/second*
Waste Collection Tray Volume
Approximately 50ml
Void Volume
500µl
*For standard Luciferase Assay or Dual-Luciferase
®
Assays, we recommend an injector speed of 200µl per
second.
Warning:
Make sure sufficient volume is available in the well to accept the injected volume, especially when
injectors are used within a loop function. Failure to do so may cause an overflow of reagent and/or sample into
adjacent wells or into the interior of the instrument.
The Injectors may also be accessed from the
Tools
icon in the menu bar.
See Setting Up and Operating the Injector System (Section 4 for more details).
!