Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
37
www.promega.com
TM470 · 4/16
4.A. Priming Injectors
1.
Place the reagent container (tube or bottle) into the reagent holder.
2.
Cover the bottle opening with Parafilm
®
to prevent reagent evaporation.
3.
Insert the injector tube through the Parafilm
®
and into the reagent bottle.
4.
Ensure the Waste Collection Tray is in position and empty.
5.
Tap the
Start
button adjacent to the Prime Injectors information in Figure 50.
13626T
A
Figure 51. Injector prime function.
6.
Indicate the injector to prime. Set the volume.
Note:
The minimum volume needed to prime is 500µl.
7.
Tap the Start Prime button (Figure 51).
8.
After priming is complete, empty the Waste Collection Tray.
Once the injectors are primed, they are ready for use according to the programmed protocol.