8 Create a new Method
22
8.1.2 Checkweigher
With the aid of the «Checkweigher», you can weigh the sample exactly to check that it matches a defined
reference value plus/minus permissible deviations. "
+
", "
–
" and "
!
II
"
" are active in the display.
If "
!
II
"
" lights up, this indicates that the measurement lies within the defined tolerances, and the drying pro-
cess can commence. If the sample weight lies outside the starting weight tolerance, the drying process can-
not be activated. In this case the starting weight tolerances are displayed on the screen as an error message.
:
CheckWeigher:
On,
Off
CheckWeigher enabled/disabled
Nominal: 100.000 g
Nominal weight
TO: 2.5% (or 0.000 g)
Over limit according to the application as weight in the basic unit, in percent
TU: 2.5% (or 0.000 g)
Under limit according to the application as weight in the basic unit, in percent
Limit type: Percent
, Weight
Limits TO/TU according to the application as weight in the basic unit, in percent
Acoustic signal: On
, Off
Acoustic signal according the weight in relation to the nominal value and the limits
8.2
Edit an Existing Method/ Create a new Method by Editing an Existing
one
Switch-on the moisture analyser:
Press
«ON/OFF»
Select the method:
• Press the soft key
«Method»
• Keep pressing
{
!
}
until the method group of your choice is selected
• Press
{
E
}
to confirm your selection
Method group - DAIRY- New Method
CheckWeigher
On
Nominal
2
TO
2.2
TU
1.8
Limit type
Weight
Acoustic signal
#
On
ESC
"
!
E
Method group
#
Corn (5)
Dairy (3)
Juice (6)
NEW
ESC
DEL
Edit
"
!
E